Aniosyme Prime
Pre-treatment disinfectant detergent foam for medical and surgical instruments.
Product Details
- Formulation combining two patented concepts of efficiency
- Prevents soil from drying
- Antimicrobial efficacy from 15 min
- Dense foam: instruments are covered evenly
- Wide compatibility
Application area
Pre-treatment of surgical instrumentation in foam: during a dry mode of transport or during storage of soiled instruments before machine or manual washing.
Maintains an adequate level of humidity to avoid blood and other organic materials drying.
Instructions for use
Ready-to-use solution.
- Beforehand, spray the bottom of the soaking tray with Aniosyme Prime and place instruments collected.
- Re-cover all the instruments with foam. Minimum contact time in the system: 15 minutes. For tuberculocidal activity, allow 30 minutes contact time.
- Cover the tray with the lid.
- Thoroughly rinse treated medical devices with tap water to remove all traces of product and organic soiling before machine or manual cleaning.
Not suitable for instruments made of zinc or iron.
Ingredients
Quaternary ammonium carbonate, non-ionic surfactants, enzymatic complex (protease, lipase, amylase), sequestering agents, excipients.
Storage
Dangerous – respect the precautions for use (drawn up according to the European rules in force regarding the classification and labelling of chemical products).
Storage: between +5°C and +35°C.
Microbiological properties
CLAIM/CONDITIONS | CONTACT TIME |
---|---|
Bactericidal According to EN 13727 and EN 14561, dirty conditions |
15 min |
Yeasticidal According to EN 13624 and EN 14562, dirty conditions |
15 min |
Tuberculocidal According to EN 14348 and EN 14563, dirty conditions |
30 min |
Mycobactericidal According to EN 14348 and EN 14563, dirty conditions |
60 min |
Effective against enveloped viruses According to EN 14476 / EN 17111, dirty conditions |
15 min |
Medical device.
Please read label and product information.
For professional use only.
0459